CVE Published: 14/11/2024 |
CVE Updated: 18/11/2024 |
CVE Year: 2024 Source: Baxter |
Vendor: Baxter |
Product: Life2000 Ventilation System Status : PUBLISHED
CVE-2024-9832 Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Metrics
CVSS Version: 3.1 |
Base Score: 9.3 CRITICAL Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H