CVE-2024-9832 Vulnerability Details

  /     /     /  

CVE-2024-9832 Metadata Quick Info

CVE Published: 14/11/2024 | CVE Updated: 18/11/2024 | CVE Year: 2024
Source: Baxter | Vendor: Baxter | Product: Life2000 Ventilation System
Status : PUBLISHED

CVE-2024-9832 Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Metrics

CVSS Version: 3.1 | Base Score: 9.3 CRITICAL
Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* LOCAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* CHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* HIGH
    Integrity Impact (I)* HIGH
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-307
CWE Name: CWE-307 Improper Restriction of Excessive Authentication Attempts
Source: Baxter

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID: CAPEC-49