CVE-2024-48974 Vulnerability Details

  /     /     /  

CVE-2024-48974 Metadata Quick Info

CVE Published: 14/11/2024 | CVE Updated: 18/11/2024 | CVE Year: 2024
Source: Baxter | Vendor: Baxter | Product: Life2000 Ventilation System
Status : PUBLISHED

CVE-2024-48974 Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device\'s configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

Metrics

CVSS Version: 3.1 | Base Score: 9.3 CRITICAL
Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* LOCAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* CHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* HIGH
    Integrity Impact (I)* HIGH
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-494
CWE Name: CWE-494 Download of Code Without Integrity Check
Source: Baxter

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID: CAPEC-186
CAPEC Description: CAPEC-186 Malicious Software Update