CVE-2024-48973 Vulnerability Details

  /     /     /  

CVE-2024-48973 Metadata Quick Info

CVE Published: 14/11/2024 | CVE Updated: 18/11/2024 | CVE Year: 2024
Source: Baxter | Vendor: Baxter | Product: Life2000 Ventilation System
Status : PUBLISHED

CVE-2024-48973 Description

The debug port on the ventilator\'s serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Metrics

CVSS Version: 3.1 | Base Score: 9.3 CRITICAL
Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* LOCAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* CHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* HIGH
    Integrity Impact (I)* HIGH
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-1263
CWE Name: CWE-1263 Improper Physical Access Control
Source: Baxter

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID: CAPEC-117
CAPEC Description: CAPEC-117 Interception


Source: NVD (National Vulnerability Database).