CVE-2023-25931 Vulnerability Details

  /     /     /  

CVE-2023-25931 Metadata Quick Info

CVE Published: 01/03/2023 | CVE Updated: 02/08/2024 | CVE Year: 2023
Source: Medtronic | Vendor: Medtronic | Product: InsterStim Applications
Status : PUBLISHED

CVE-2023-25931 Description

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Metrics

CVSS Version: 3.1 | Base Score: 6.4 MEDIUM
Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* PHYSICAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* LOW
    Integrity Impact (I)* HIGH
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-620
CWE Name: CWE-620: Unverified Password Change
Source: Medtronic

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).