CVE-2023-0888 Vulnerability Details

  /     /     /  

CVE-2023-0888 Metadata Quick Info

CVE Published: 13/03/2023 | CVE Updated: 02/08/2024 | CVE Year: 2023
Source: B.Braun | Vendor: B. Braun | Product: Space Battery Pack SP with Wi-Fi
Status : PUBLISHED

CVE-2023-0888 Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Metrics

CVSS Version: 3.1 | Base Score: 4.9 MEDIUM
Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N

l➤ Exploitability Metrics:
    Attack Vector (AV)* NETWORK
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* HIGH
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* NONE
    Integrity Impact (I)* HIGH
    Availability Impact (A)* NONE

Weakness Enumeration (CWE)

CWE-ID: CWE-95
CWE Name: CWE-95 Eval injection
Source: B. Braun

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID: CAPEC-233
CAPEC Description: CAPEC-233 Privilege Escalation


Source: NVD (National Vulnerability Database).