CVE-2022-38100 Vulnerability Details

  /     /     /  

CVE-2022-38100 Metadata Quick Info

CVE Published: 13/09/2022 | CVE Updated: 16/09/2024 | CVE Year: 2022
Source: icscert | Vendor: Contec Health | Product: CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Status : PUBLISHED

CVE-2022-38100 Description

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

Metrics

CVSS Version: 3.1 | Base Score: 7.5 HIGH
Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* NETWORK
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* NONE
    Integrity Impact (I)* NONE
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-400
CWE Name: CWE-400 Uncontrolled Resource Consumption
Source: Contec Health

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).