CVE Published: 13/09/2022 |
CVE Updated: 16/09/2024 |
CVE Year: 2022 Source: icscert |
Vendor: Contec Health |
Product: CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor Status : PUBLISHED
CVE-2022-38100 Description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Metrics
CVSS Version: 3.1 |
Base Score: 7.5 HIGH Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H