CVE-2022-36385 Vulnerability Details

  /     /     /  

CVE-2022-36385 Metadata Quick Info

CVE Published: 13/09/2022 | CVE Updated: 16/09/2024 | CVE Year: 2022
Source: icscert | Vendor: Contec Health | Product: CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Status : PUBLISHED

CVE-2022-36385 Description

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.

Metrics

CVSS Version: 3.1 | Base Score: 6.8 MEDIUM
Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

l➤ Exploitability Metrics:
    Attack Vector (AV)* PHYSICAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* HIGH
    Integrity Impact (I)* HIGH
    Availability Impact (A)* HIGH

Weakness Enumeration (CWE)

CWE-ID: CWE-284
CWE Name: CWE-284 Improper Access Control
Source: Contec Health

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).