CVE-2021-33886 Description

An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.

Metrics

CVSS Version: 3.1 | Base Score: 8.1 HIGH
Vector: CVSS:3.1/AC:L/AV:A/A:N/C:H/I:H/PR:N/S:U/UI:N

l➤ Exploitability Metrics:
    Attack Vector (AV)* ADJACENT_NETWORK
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* HIGH
    Integrity Impact (I)* HIGH
    Availability Impact (A)* NONE

Weakness Enumeration (CWE)

CWE-ID:
CWE Name: n/a
Source: n/a

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).