CVE-2020-25166 Description

An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.

Metrics

CVSS Version: 3.1 | Base Score: 7.6 HIGH
Vector: CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L

l➤ Exploitability Metrics:
    Attack Vector (AV)* NETWORK
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* LOW
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* LOW
    Integrity Impact (I)* HIGH
    Availability Impact (A)* LOW

Weakness Enumeration (CWE)

CWE-ID: CWE-347
CWE Name: CWE-347: Improper Verification of Cryptographic Signature