CVE-2020-10627 Vulnerability Details

  /     /     /  

CVE-2020-10627 Metadata Quick Info

CVE Published: 01/12/2021 | CVE Updated: 04/08/2024 | CVE Year: 2020
Source: icscert | Vendor: Insulet | Product: Omnipod Insulin Management System
Status : PUBLISHED

CVE-2020-10627 Description

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Metrics

CVSS Version: 3.1 | Base Score: 7.3 HIGH
Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L

l➤ Exploitability Metrics:
    Attack Vector (AV)* LOCAL
    Attack Complexity (AC)* LOW
    Privileges Required (PR)* NONE
    User Interaction (UI)* NONE
    Scope (S)* UNCHANGED

l➤ Impact Metrics:
    Confidentiality Impact (C)* LOW
    Integrity Impact (I)* HIGH
    Availability Impact (A)* LOW

Weakness Enumeration (CWE)

CWE-ID: CWE-284
CWE Name: CWE-284 Improper Access Control
Source: Insulet

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).