CVE-2018-8868 Vulnerability Details

  /     /     /  

CVE-2018-8868 Metadata Quick Info

CVE Published: 02/07/2018 | CVE Updated: 17/09/2024 | CVE Year: 2018
Source: icscert | Vendor: ICS-CERT | Product: Medtronic MyCareLink Patient Monitor
Status : PUBLISHED

CVE-2018-8868 Description

Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor\'s communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Metrics

CVSS Version: 3.1 | Base Score: n/a
Vector: n/a

l➤ Exploitability Metrics:
    Attack Vector (AV)*
    Attack Complexity (AC)*
    Privileges Required (PR)*
    User Interaction (UI)*
    Scope (S)*

l➤ Impact Metrics:
    Confidentiality Impact (C)*
    Integrity Impact (I)*
    Availability Impact (A)*

Weakness Enumeration (CWE)

CWE-ID: CWE-749
CWE Name: EXPOSED DANGEROUS METHOD OR FUNCTION CWE-749
Source: ICS-CERT

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).