CVE-2018-14781 Vulnerability Details

  /     /     /  

CVE-2018-14781 Metadata Quick Info

CVE Published: 13/08/2018 | CVE Updated: 17/09/2024 | CVE Year: 2018
Source: icscert | Vendor: ICS-CERT | Product: Medtronic insulin pump
Status : PUBLISHED

CVE-2018-14781 Description

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Metrics

CVSS Version: 3.1 | Base Score: n/a
Vector: n/a

l➤ Exploitability Metrics:
    Attack Vector (AV)*
    Attack Complexity (AC)*
    Privileges Required (PR)*
    User Interaction (UI)*
    Scope (S)*

l➤ Impact Metrics:
    Confidentiality Impact (C)*
    Integrity Impact (I)*
    Availability Impact (A)*

Weakness Enumeration (CWE)

CWE-ID: CWE-294
CWE Name: AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294
Source: ICS-CERT

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).