CVE-2014-5406 Vulnerability Details

  /     /     /  

CVE-2014-5406 Metadata Quick Info

CVE Published: 06/07/2015 | CVE Updated: 06/08/2024 | CVE Year: 2014
Source: icscert | Vendor: n/a | Product: n/a
Status : PUBLISHED

CVE-2014-5406 Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Metrics

CVSS Version: 3.1 | Base Score: n/a
Vector: n/a

l➤ Exploitability Metrics:
    Attack Vector (AV)*
    Attack Complexity (AC)*
    Privileges Required (PR)*
    User Interaction (UI)*
    Scope (S)*

l➤ Impact Metrics:
    Confidentiality Impact (C)*
    Integrity Impact (I)*
    Availability Impact (A)*

Weakness Enumeration (CWE)

CWE-ID:
CWE Name: n/a
Source: n/a

Common Attack Pattern Enumeration and Classification (CAPEC)

CAPEC-ID:
CAPEC Description:


Source: NVD (National Vulnerability Database).